Informações de encomenda

ECGInput dynamic range±(0.5mVp~5mVp)
Differential input impedance≥10MΩ
Bandwidth0.05~150Hz (Diagnostic)
0.5~40Hz (Monitoring)
1~20Hz (Operation)
CMRR≥90dB (Diagnostic)
≥105dB (Monitoring & Operation)
Sensitivity selection×1/4, ×1/2, ×1, ×2, ×4 and Auto
Sweeping speed6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
HR measuring range15~350bpm
HR accuracy±1% or ±2bpm, whichever is greater
Pacemaker pulse detection and rejection function
RESPMeasuring range0~120rpm
Measuring accuracy±5% or ±2 rpm, whichever is greater
TEMPMeasuring range21.0~50.0℃
Measuring accuracy±0.2℃ from 25~45℃
NIBPTechniqueOscillometric method
Typical measurement time<30 seconds (adult cuff)
NIBP measuring range

SYS: 25-290mmHg (Adult)

25-240mmHg (Pediatric)

25-140mmHg (Neonate)

NIBP measuring range

DIA: 10-250mmHg (Adult)

10-200mmHg (Pediatric)

10-115mmHg (Neonate)

NIBP measuring range

MAP:15-260mmHg (Adult)

15-215mmHg (Pediatric)

15-125mmHg (Neonate)

NIBP measuring accuracyMean difference: ±5mmHg

Standard deviation: 8mmHg
NIBP measurement modeManual, Auto, STAT, Multi-cycle mode
Auto measuring intervals1-480min
SpO2TechniqueDual-wavelength optical method
Measuring range0%~100%
Measuring accuracyArms are not greater than 2% for SpO2 range 70~100%.
PR measuring range30~250bpm
PR measuring accuracy±2bpm or ±2%, whichever is greater
Low perfusion performanceAs low as 0.3%.
CO2TechniqueInfrared optical method
Sampling modeSidestream or Mainstream
Measuring range0~150mmHg
Measuring accuracy:0~40mmHg ±2mmHg
41~70mmHg ±5% of reading
71~100mmHg ±8% of reading
101~150mmHg ±10% of reading
Flow rate50ml/min ±10 ml/min (Sidestream)
Cerebral State Monitoring (CSM)EEG sensitivity±400µV
Noise level<2µVp-p, <0.4µV RMS (1~250Hz)
CMRR>140dB
Input impedance>50Mohm
CSI and update0-100. filter: 6-42Hz, 1 sec. update
EMG%0-100 (logarithmic)
filter: 75-85 Hz, 1 sec. update
BS%0-100. filter: 2-42 Hz, 1 sec. update
IBPTechniqueStrain gauge transducer

Input sensitivity5µV/V/mmHg

Measuring range-50~300mmHg

Measuring accuracy±2% or ±4mmHg, whichever is greater

Measuring positionsART, RAP, PA, LAP, CVP

ICP, AUXP1, AUXP2

Calibrationzero calibrating
Cardiac Output (C.O.)Blood temperature23-43℃, accuracy: ±0.5℃
measuring: range
Injecta temperature0-20℃, accuracy: ±0.5℃
measuring: range
Measuring range0.2~20 L/min
Measuring accuracy±0.2 L/min or ±10%, whichever is greater
Other SpecificationsPower supplyAC 100V-240V, 50/60Hz, 60VA
Built-in lithium battery11.1V/4400mAh
Display12.1 inch TFT display
Alarming method3 levels audible-visible alarm
NetworkingEthernet
Standard configurationECG, Respiration, SpO2, PR, NIBP, Temperature
OptionTouch Screen, 2-IBP, EtCO2, Nellcor SpO2, SunTech NIBP, 12-lead ECG, Cardiac Output, Cerebral